| Parametro | Valore |
|---|---|
| Articolo | Specifications |
| Peso | Net weight: 300 g approx. |
| Dettagli | Controller |
| Prodotto | Ultra High-speed Laser Displacement Sensor [CCD Style] |
| Accessorio | - |
| Numero di parte | HL-C1C-M |
| Uscita allarme | Photo-MOS relay- Maximum load current: 50 mA- Applied voltage: 30 V DC or less (between alarm output and COM)- ON impedance: 35 Ohm or less- Operation time: Max. 2 ms |
| Lunghezza del cavo | Power line: Less than 10 m 32.808 ftSignal line: Less than 30 m 98.425 ft |
| Nome del prodotto | Ultra High-speed Laser Displacement Sensor HL-C1(Discontinued) |
| Frequenza di campionamento | Selectable from 100 micro s / 144 micro s / 200 micro s / 255 micro s / 332 micro s / 498 micro s / 1,000 micro s |
| Funzioni di attesa | Selectable from NORMAL / P-P / PEAK / VALLEY(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Numero di prodotto | HL-C1C-M |
| Tensione di alimentazione | 24 V DC plus or minus 10 % including ripple 0.5 V (P-P) |
| Filtri digitali | High pass: OFF, 10 to 2,000 Hz (9 steps)Low pass: OFF, 10 to 2,000 Hz (9 steps)(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Indicatori: SCURO | Red LED (lights up upon disabled measurement due to insufficient light at sensor head 1 or 2) |
| Ingresso azzerato ON | Zero set: ON when voltage (using input voltage: 12 to 24 V DC, maximum input voltage: 30 V DC) is input |
| Calibrazione: Span | 0.9000 to 1.1000(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Ingresso di azzeramento OFF | Zero set: OFF when voltage (using input voltage: 12 to 24 V DC, maximum input voltage: 30 V DC) is input |
| Calibrazione: Spostamento | plus or minus 20.0000 mm plus or minus 0.787 in(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Consumo attuale | When 1 sensor is connected: 430 mA approx.When 2 sensors are connected: 550 mA approx. |
| Indicatori: LUMINOSI | Red LED (lights up upon disabled measurement due to excessive light at sensor head 1 or 2) |
| Funzioni di calcolo | L plus or minus KA, L plus or minus KB, L plus or minus K (A plus or minus B)A, B: Sensor head 1, Sensor head 2 measurement values, L = plus or minus 999.9999, K = 0.0001 to 99.9999(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Ingresso/uscita seriale | RS-232C |
| Testa del sensore Connectale | Maximum 2 sensor heads |
| Uscita analogica: corrente | Output current: 4 to 20 mA/F.S. [default setting when diffuse reflective mode is selected (Note 2)]Output range: 0 to 29.5 mA (maximum of 25 mA at max. load impedance)Load impedance: 250 Ohm or less(Note 1) The maximum analog output current will vary with load impedance.(Note 2) If specular reflective mode is selected, then the default setting is 5.6 to 18.4 mA/F.S. |
| Uscita analogica: tensione | Output voltage: plus or minus 5 V/F.S. [default setting when diffuse reflective mode is selected (Note) ]Output range: -10.9 to +10.9 VOutput current: Max. 2 mA, Output impedance: 50 Ohm(Note) If specular reflective mode is selected, then the default setting is plus or minus 4 V/F.S. |
| Numero medio di campioni | OFF, 2 to 32,768 times (16 steps)(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Risultati del giudizio (O1, O2) | Photo-MOS relay- Maximum load current: 50 mA- Applied voltage: 30 V DC or less (between judgment output and COM)- ON impedance: 35 Ohm or less- Operation time: Max. 2 ms |
| Indicatori: Emissione laser | Green LED (lights up during laser emission from sensor head 1 or sensor head 2, or immediately before laser emission) |
| Caratteristiche di temperatura | plus or minus 0.01 % F.S./degrees |
| Input di temporizzazione (emissione laser) | Laser emission stops or continues when voltage (using input voltage: 12 to 24 V DC, maximum input voltage: 30 V DC) is input or there is an open circuit: determined based on input mode selection. |
| Uscita allarme: funzionamento dell'uscita | Opened when the amount of light is excessive or insufficient. |
| Uscita allarme: protezione da cortocircuito | Incorporated |
| Resistenza ambientale: Umidità ambientale | 35 to 85 % RH, Storage: 35 to 85 % RH |
| Resistenza ambientale: Resistenza agli urti | 196 m/s2 (20 G approx.) in X, Y and Z directions for 3 times each |
| Uscite di giudizio (O1, O2): operazione di uscita | Opened or closed when the threshold value is reached. Determined based on judgment output mode selection. (The threshold value varies with the hysteresis setting.) |
| Resistenza ambientale: Temperatura ambiente | 0 to +50 degrees +32 to +122 degrees Fahrenheit (No dew condensation), Storage: -20 to +70 degrees -4 to +158 degrees Fahrenheit |
| Resistenza ambientale: Resistenza alle vibrazioni | 10 to 55 Hz frequency (period: 1 min.) 0.75 mm 0.030 in amplitude in X,Y and Z directions for 30 min. each |
| Risultati del giudizio (O1, O2): Categoria di utilizzo | DC-12 or DC-13 |
| Uscite di giudizio (O1, O2): protezione da cortocircuito | Incorporated |